Abstract

To establish a simple method to determine breast readout accuracy on mammography units. A thickness measuring device (TMD) was used in conjunction with a breast phantom. This phantom had compression characteristics similar to human female breast tissue. The phantom was compressed, and the thickness was measured using TMD and mammography unit readout. Measurements were performed on a range of screen film mammography (SFM) and full-field digital mammography (FFDM) units (8 units in total; 6 different models/manufacturers) for two different sized paddles and two different compression forces (60 and 100 N). The difference between machine readout and TMD for the breast area, when applying 100 N compression force, for nonflexible paddles was largest for GE Senographe DMR+ (24 cm × 30 cm paddle: +14.3%). For flexible paddles the largest difference occurred for Hologic Lorad Selenia (18 cm × 24 cm paddle: +26.0%). None of the units assessed were found to have perfect correlation between measured and readout thickness. TMD measures and thickness readouts were different for the duplicate units from two different models/manufacturers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.