Abstract

The authors describe a new radionuclide method for assessing the functional integrity of slow intrathecal infusion devices. Approximately 11 mCi (400 MBq) of Tc-99m DTPA is injected into the pump chamber. Early and delayed imaging of the pump, tubing, and spinal cerebrospinal fluid space is performed on a digital large-field-of-view gamma camera. The digitally displayed images are then reviewed by adjustment of the grey scale window. Four patients have had eight studies and in each case the result has been confirmed by surgical exploration or by clinical response to change in therapy.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.