Abstract

In this work, an experimental study of the radiation beam characteristics of the photon beams produced by commercial irradiator that uses a single collimated 137Cs source is performed. A set of lead attenuators are placed at the exit window of the irradiator to vary the air Kerma that is required to cover the instrument scales at a particular calibration distance. The results strongly enhance the findings of previous studies that a single source with appropriate attenuation and collimation is a practical solution for calibration laboratories with an acceptable overall accuracy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.