Abstract

The relationship between International Organization for Standardization (ISO) accuracy criteria and mean absolute relative difference (MARD), 2 methods for assessing the accuracy of blood glucose meters, is complex. While lower MARD values are generally better than higher MARD values, it is not possible to define a particular MARD value that ensures a blood glucose meter will satisfy the ISO accuracy criteria. The MARD value that ensures passing the ISO accuracy test can be described only as a probabilistic range. In this work, a Bayesian model is presented to represent the relationship between ISO accuracy criteria and MARD. Under the assumptions made in this work, there is nearly a 100% chance of satisfying ISO 15197:2013 accuracy requirements if the MARD value is between 3.25% and 5.25%.