Abstract

The optimum slice location within the pulmonary root to quantify pulmonary valve haemodynamics was examined using magnetic resonance (MR) phase velocity mapping. MRI was carried out on 15 patients with congenital aortic valve disease. Although the patients had aortic valve disease, all measurements were made on the pulmonary valve. Systolic (Q(SYS)) and diastolic (Q(DIAS)) blood flow volume and cardiac index (CI) were determined at four pulmonary artery locations. The change in diastolic flow volume relative to slice 1, closest to the pulmonary valve, was also calculated. For a change in axial position of 1.5 cm, i.e. from 0.5 to 2 cm from the annulus, there was a change in diastolic flow volume of 4.4 ml. There was a significant increase in the mean diastolic flow from 3.4 to 7.7 ml (p = 0.01 between slice positions 0.5 and 2 cm. However, there was no significant change in CI, 3.4-3.7 l/min/m2 (p = 0.14) over the same distance. We believe that two factors are responsible for these results. The first is that of compliance, whose effects can be minimized by placing the MR slice close to the valve, however, this will not account for the second factor, being that of valve motion, and hence diastolic pulmonary valve flow or regurgitant volume will be underestimated. The degree of underestimation may only be important at mild and moderate levels of regurgitation or if changes in regurgitation are to be temporally measured.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.