Abstract

A new project is presented, the pumping oxygenator, functionally integrating pulsatile pumping and blood oxygenation in a single device. Solid, semipermeable silicone membranes allow gas exchange and simultaneously transfer energy from pressurized gas to blood thanks to their distensibility and to inlet and outlet 1-way valves. Two small-sized (1 m2 exchange surface area) prototypes were designed, constructed, hydraulically characterized, and subjected to gas transfer evaluation tests. Blood flow rates (Q(b)) up to 1,250 ml/min were obtained with 30 mm Hg static preload and 130 mm Hg afterload with 0.7 m upstream and 2.1 m downstream 3/8 inch pipes. Physiological oxygen transfer (VO2 = 5 ml/dl, ml of transferred O2/dl of treated blood) was delivered at Q(b) < 900 ml/min, about 4 ml/dl at Q(b) = 1,250 ml/min. VO2 also was significantly increased by increasing percent systolic time. CO2 transfer decreased regularly with increasing Q(b) from VCO2 = 4.8 ml/dl at Q(b) = 400 ml/min to VCO 2 = 2.1 ml/dl at Q(b) = 1,250 ml/min. The results confirm the possibility of integrating oxygenation and pulsatile pumping. The pumping oxygenator represents a promising project deserving further improvements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.