Abstract

Pressures in the right side of the heart and pulmonary capillary wedge can be obtained by cardiac catheterization without the aid of fluoroscopy. A No. 5 French double-lumen catheter with a balloon just proximal to the tip is inserted into the right atrium under pressure monitoring. The balloon is then inflated with 0.8 ml of air. The balloon is carried by blood flow through the right side of the heart into the smaller radicles of the pulmonary artery. In this position when the balloon is inflated wedge pressure is obtained. The average time for passage of the catheter from the right atrium to the pulmonary artery was 35 s in the first 100 passages. The frequency of premature beats was minimal, and no other arrhythmias occurred.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.