Abstract

The study aims to measure the strength of the proximal fixation of endografts in short and long necks. Three types of endografts were compared: Gore Excluder, Vascutek Anaconda and Medtronic Endurant. The proximal part of the stent grafts was inserted in bovine arteries and the graft was then attached to a tensile testing machine. The force to obtain dislodgement (DF) from the aorta was recorded for each graft at proximal seal lengths of 10 and 15 mm. The median DF (interquartile range, IQR) for the Excluder, the Anaconda and the Endurant with a seal length of 15 mm was: 11.8 (10.5-12.0) N, 20.8 (18.0-30.1) N and 10.7 (10.4-11.3) N. With the shorter proximal seal of 10mm, DF was, respectively: 6.0 (4.5-6.6) N, 17.0 (11.2-36.6) N and 6.4 (6.1-12.0) N. The proximal fixation of the Anaconda is superior to the Excluder and the Endurant at short necks of 10 and 15 mm in an experimental set-up. There is a statistically significant decrease of proximal fixation for the Excluder stent graft, when decreasing the length of the proximal neck from 15 to 10 mm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.