Abstract

The administration of a uniform dose rate to the vaginal surface is important in the management of endometrial, cervical, and vaginal malignancies. Unfortunately, conventional vaginal cylinders fail to provide this uniformity, and although dome cylinders do so, they require specialized 137Cs sources. Thus, a new acrylic vaginal cylinder has been developed to use with standard 137Cs sources and provides a uniform dose rate independent of vaginal size. Each contoured cylinder follows a particular isodose line. A metal ring is used to secure the device to minimize vulvar trauma associated with other vaginal cylinders. The construction and use of a set of these cylinders has allowed determination of their utility and limitations, which will be discussed in detail. Although these applicators do not completely replace conventional cylinders, they do offer a useful addition to the brachytherapeutic armamentarium.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.