Abstract

The proton radiation facility is built at an irradiation terminal of a 100-MeV cyclotron accelerator in the China Institute of Atomic Energy (CIAE). It is mainly used for testing the single-event effect of electronic devices. The part and function of the proton beamline for radiation effects research are introduced. Many detectors and techniques are used to accomplish accurate and reliable dosimetry and characterization of the proton beam. The results of beam characterization such as transmission, energy, lateral profile, and flux are reported. A typical monitoring and measuring system consist of a secondary electron emission monitor (SEM) working in a transmission mode and Faraday cup (FC). A cross-checking method of proton flux accuracy by FC and scintillation detector (SC) is established. The application of the reference single-event upset (SEU) monitor for the characterization of proton beams at different facilities is described. The SEU cross-section of the Reference SEU Monitor has confirmed the flux detection accuracy at CIAE.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.