BackgroundThe Lotrach endotracheal tube has a unique low-volume, low-pressure (LVLP) cuff, which has been designed to prevent pressure injury to the tracheal wall. We aimed to estimate the pressure exerted on the tracheal wall by the LVLP cuff and a conventional cuff in a bench-top, clinical and radiological study.MethodIn the bench-top study, a model trachea was intubated with the LVLP cuff and the conventional cuff. The cuff pressure was controlled using a constant pressure device. We assessed the pressure exerted on the tracheal wall by measuring the ability of the cuffs to support a column of water using a standard protocol. In the clinical study, we tested the ability of both cuffs to prevent air leak during a staged recruitment manoeuvre. In the radiological study, we recorded the degree of anatomical distortion of the trachea from both cuffs in the antero-posterior (AP) and transverse tracheal diameters. We performed statistical analysis using non-inferiority tests.ResultsIn the bench-top study, the LVLP cuff achieved a plateau at a mean height of 25.2 cmH2O (SD 0.34). In contrast, the conventional cuff failed to maintain any water above the cuff and a plateau could not be measured. In the clinical study, the mean pressure at which air leak occurred was 30.0 +/- 0.8 cmH2O (SD 3.8) using the LVLP cuff and 32.4 +/- 0.7 cmH2O (SD 3.0) using the conventional cuff. In the radiological study, the mean degree of anatomical distortion of the trachea in AP and transverse tracheal diameter was 2.9 +/- 2.2 mm (SD 2.1) and 1.8 +/- 1.4 mm (SD 1.4) using the LVLP cuff and 4.4 +/- 1.3 mm (SD 1.4) and 2.6 +/- 1.5 mm (SD 1.6) using the conventional cuff.ConclusionsThe bench-top and clinical studies both demonstrated that the LVLP cuff exerted approximately 30 cmH2O of pressure on the tracheal wall. These results are supported by our radiological study. We conclude that the LVLP cuff exerts an acceptable amount of pressure on the tracheal wall when it is operated at the recommended intracuff pressure.

Highlights

  • The Lotrach endotracheal tube has a unique low-volume, low-pressure (LVLP) cuff, which has been designed to prevent pressure injury to the tracheal wall

  • We conclude that the LVLP cuff exerts an acceptable amount of pressure on the tracheal wall when it is operated at the recommended intracuff pressure

  • The clinical study estimated that the LVLP cuff exerted approximately 30 cmH2O of pressure on the tracheal wall, which was similar to that exerted by the HVLP cuff

Read more

Summary

Introduction

The Lotrach endotracheal tube has a unique low-volume, low-pressure (LVLP) cuff, which has been designed to prevent pressure injury to the tracheal wall. We aimed to estimate the pressure exerted on the tracheal wall by the LVLP cuff and a conventional cuff in a bench-top, clinical and radiological study. It is important to ensure an appropriate amount of pressure is exerted on the tracheal wall by an endotracheal tube cuff for two opposing reasons. A Traditional endotracheal tube cuffs are high-volume, low-pressure (HVLP) cuffs. These cuffs should not be fully inflated when used. The pressure exerted on the tracheal wall by the cuff is equal to the intracuff pressure. At an intracuff pressure of 30 cmH2O, the HVLP cuff exerts approximately 30 cmH2O of pressure on the tracheal wall. It is recommended that HVLP cuffs are operated at an intracuff pressure of 30 cmH2O

Objectives
Methods
Results
Discussion
Conclusion