Abstract

To assess the precision of a new spectral-domain optical coherence tomographer (SD-OCT)/Placido topographer (MS-39; CSO) and its comparison with a swept-source OCT (SS-OCT) biometer (Argos; Movu, Inc) in patients with cataract. Fifty-three right eyes from 53 patients were examined by two experienced operators three times using both devices randomly. Employing the within-subject standard deviation (Sw), test-retest variability, coefficient of variation, and intraclass correlation coefficient to evaluate intraoperator repeatability and interoperator reproducibility; the double-angle plots to analyze astigmatism; and Bland-Altman plots and 95% limits of agreement to verify the agreement between devices. The SD-OCT/Placido tomographer showed high precision, with coefficient of variation of 0.44% or less, intraclass correlation coefficient of 0.945 or greater for all parameters, test-retest variability of 4.21 µm or less for central corneal thickness (CCT), 0.03 mm or less for anterior chamber depth (ACD) and aqueous depth (AQD), and 0.25 diopters (D) or less for mean keratometry (Km), J0, and J45. The inter-device differences in Km, J0, and J45 were statistically insignificant, whereas the remaining were statistically but not clinically significant. The 95% limits of agreement of CCT, ACD, AQD, Km, J0, and J45 were -3.70 to 15.25 µm, -0.06 to 0.04 mm, -0.06 to 0.04 mm, -0.28 to 0.35 D, -0.27 to 0.26 D, and -0.27 to 0.21 D, respectively. The double-angle plot confirmed the high agreement in astigmatism. For CCT, ACD, AQD, Km, and astigmatism measurements in patients with cataract, the new SD-OCT/Placido tomographer has excellent precision and high agreement with the Argos SS-OCT biometer, and can be used interchangeably. [J Refract Surg. 2021;37(9):616-622.].

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call