Abstract

By refining the technique and analysis of an earlier paper, it is shown that a compensation method is the basis of the most precise measurement of noise temperature, especially for cryogenic noise generators. The measurement system is adjusted so that it is in thermal equilibrium with the generator under test. As is typical for the compensation method, the generator under test may be mismatched, and the comparison errors and detector limitations can be verified experimentally. For generators with small reflection coefficients (|/spl Gamma/|is<or =0.1), the comparison error is very close to the theoretical minimum, namely, twice the system resolution. For reflection coefficients as large as 0.5, the comparison error increases in a typical system to about four times the system resolution.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.