Abstract

Oscillometric devices in contrast to the traditional Doppler based method for ankle-brachial index measurements have promising advantages like no need for special training, faster performance, and operator independence. Comparative assessment of the oscillometric and Doppler-based ankle-brachial index measurement. Ankle-brachial index measurements were performed by continuous wave Doppler and an automatic oscillometric device (BOSO ABI-system 100) in consecutive subjects. The comparative assessment was performed by Bland-Altman and ROC analysis. The two kinds of measurements (734 measurements) showed a good agreement in the ankle-brachial index spectrum close to the cut-off value of 0.9. The agreement diminished below or above this value. The optimal oscillometric ankle-brachial index diagnostic cut-off value was 0.96. The oscillometric device is not interchangeable for Doppler devices in the whole ankle-brachial index spectrum. Nevertheless, owing to its discriminative power, the oscillometric measurement potentially has an efficient role in the screening of asymptomatic patients. Orv Hetil. 2018; 159(5): 176-182.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.