Abstract

In this paper we investigate the potential of alkyl-bonded silica monolithic columns for the isolation and identification of drug-related components in biological fluids. Up to 6 columns have been connected in series to produce a chromatographic system with up to 40,000 plates. This high-resolution chromatography system has been coupled to both MS and NMR to enable efficient detection and characterisation of drug-related components in biological fluids. The use of six coupled columns has been shown to give enhanced resolution over a high quality silica particulate column packed with 3 μm material which exhibits the same back pressure. The effect of volume and mass load on the performance of monolithic columns for semi-preparative chromatography of biological fluids has also been investigated. In these studies it was possible to inject up to 100 mL of neat urine with no loss of chromatographic performance. Furthermore, upon re-testing, the columns showed similar chromatographic performance. Again several columns were serially connected, producing enhanced resolution in the semi-preparative mode.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call