Abstract

The dosimetric capabilities of tissue equivalent rod-shaped polymer pencil lead graphite (PPLG) has been investigated using an Elekta HD Linac, evaluated for electrons from 6 to 15 MeV and doses in the range 0.5 Gy up to 20 Gy. These encompass the major irradiation value ranges used in electron radiation therapy procedures. Parameters concerning thermoluminescence (TL) dosimetry have been investigated, including reproducibility, TL glow curve, the linearity of dose response, sensitivity, energy response and fading. The smaller diameter PPLG samples (0.3 mm) offer excellent dosimetric characteristics, superior to that of the 0.5 mm, 0.7 mm, and 0.9 mm diameter rods. This study confirms that low-cost Pencil Lead-Graphite (PLG) dosimeters provide energy independent dose sensitive linear response over a wide dose range. The small physical size offers unchallenged point detector spatial resolution, also enjoying minimal rates of fading and insensitivity to room lighting. As such, PPLG offers a viable alternative to current commercial dosimeters, well suited to applications in radiotherapy.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.