Abstract

Ongoing uncertainties have existed regarding possible effects at low photon energies of two copper filters covering Element 1 of the Harshaw Type-8814 thermoluminescent dosimeter badge casing. To address these, Brookhaven National Laboratory's (BNL) Personnel Monitoring Group conducted a test in which Type-8814 badges with one copper filter covering Element 1 were irradiated at several low-energy levels side-by-side with the same number of badges with two copper filters covering Element 1. A review of exposures to personnel at Brookhaven Laboratory to possible low-energy photon flux was also conducted. From both the test and the review of exposures, it can be concluded that, for radiological work under the conditions at BNL, there is no apparent dosimetric difference if one or two copper filters cover Element 1 of the Type-8814 badge.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.