Abstract
Validation of process of medical product sterilization includes absorbed dose mapping in a phantom made of material representative of the object to be processed. Commonly, such measurements are carried out using the disposable chemical dosimeters placed in the phantom at the nodes of the 3D grid. Such a procedure is very laborious and costly in terms of the consumption of dosimeters. In the work, we investigated the possibility of using pyrometric method for prompt mapping of the absorbed dose. The studies were carried out using a rectangular phantom in the form of a set of expanded polystyrene plates, which is exposed to a scanned electron beam. The temperature and absorbed dose distributions in the phantom were measured. A linear dependence between them has been established. The calculation of the absorbed dose profile was also performed by MC simulations. Satisfactory agreement of the calculated dose distribution with the measured one is shown. The limitations of applicability of the proposed method are determined.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.