Abstract

The Barrel CRID detector has been operating successfully at SLD for the past seven years. It is an important tool for SLD physics analyses. We report results based on long term operational experience of a number of important quantities such as the Cherenkov quality factor, N/sub 0/ of the device, fluid transparency, electron lifetime, single electron detection efficiency, anode wire ageing, TMAE purity, long term transparency, Cherenkov angle resolution and the number of photoelectrons observed per ring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.