Abstract

This paper reports on the performance evaluation of a novel latching-type electromagnetic actuator which is designed to be embedded at selected joints along single-port laparoscopic surgical instruments (SLS). The aim of this actuator is to allow these instruments to become articulated with a push of a button in order to provide the optimum angulation required during SLS operations. This new actuator is comprised of electromagnetic coil elements, soft magnetic frames and a permanent magnet. Unlike conventional electromagnetic actuators, latching-type electromagnetic actuators could maintain their positions at either end of the actuation stroke without any power application requirement. In the current design, magnetic attraction forces initiated between the permanent magnet and the magnetic frame are utilised to lock the position of the actuator whilst a certain angulation position of the actuator is achieved as a result of the magnetic repulsion forces established between the permanent magnet and the coil elements. The performance of the new actuator in terms of the output force, maximum angulation and patient’s safety, was evaluated experimentally and the results were found to compare well with those acquired numerically using finite element methods. This actuator was seen to exhibit sufficient actuation forces and hence, it was capable of providing adaptable angulation characteristics for SLS tools. Finally, thermal evaluation of the actuator’s operation was conducted, which was found to be within safety limits specified by clinicians.

Highlights

  • IntroductionSingle-port laparoscopic surgery (SLS), which utilises one major incision instead. H

  • Magnetic attraction forces initiated between the permanent magnet and the magnetic frame are utilised to lock the position of the actuator whilst a certain angulation position of the actuator is achieved as a result of the magnetic repulsion forces established between the permanent magnet and the coil elements

  • In order to maintain consistency in this investigation, it was decided to allow the permanent magnet to start its stroke from the top-end of the central chamber in all the tests reported

Read more

Summary

Introduction

Single-port laparoscopic surgery (SLS), which utilises one major incision instead. H. The inevitable heat buildup caused by the continuous current feed to energise the magnetic circuit of these actuators has been found in certain medical applications as a problem since it breaches the patient’s safety [17] This problem was overcome in some applications using latching-type electromagnetic actuators, which utilise a permanent magnet to lock the position of the actuator instead of the magnetic energy generated by the continuous activation of the actuator’s electromagnetic circuit [18]. El Wahed dual-coil design for position locking and actuation, respectively This new actuator is proposed to be embedded along SLS surgical instruments, which can be squeezed through a standard tocar-port, to enable the tool’s angulation that is required during SLS operations. The overall performance of the new actuator is discussed and compared with other competitive actuators

Actuator Design and Optimisation
Experimental Arrangement
Least Activation Current Test Setup
Magnetic Force Test Setup
Thermal Test Setup
Results and Discussion
Least Activation Current
Actuator Magnetic Force
Actuator Thermal Performance
Summary of Actuator Performance
Conclusions