Abstract

Background: Calcaneal quantitative ultrasound (QUS) is widely used in osteoporosis screening, but the cut-off values for risk stratification remain unclear. This study validates the performance of a calcaneal QUS device (CM-200) using dual-energy X-ray absorptiometry (DXA) as the reference and establishes a new set of cut-off values for CM-200 in identifying subjects with osteoporosis. Methods: The bone health status of Malaysians aged ≥40 years was assessed using CM-200 and DXA. Sensitivity, specificity, area under the curve (AUC) and the optimal cut-off values for risk stratification of CM-200 were determined using receiver operating characteristic (ROC) curves and Youden’s index (J). Results: From the data of 786 subjects, CM-200 (QUS T-score <−1) showed a sensitivity of 82.1% (95% CI: 77.9–85.7%), specificity of 51.5% (95% CI: 46.5–56.6%) and AUC of 0.668 (95% CI: 0.630–0.706) in identifying subjects with suboptimal bone health (DXA T-score <−1) (p < 0.001). At QUS T-score ≤−2.5, CM-200 was ineffective in identifying subjects with osteoporosis (DXA T-score ≤−2.5) (sensitivity 14.4% (95% CI: 8.1–23.0%); specificity 96.1% (95% CI: 94.4–97.4%); AUC 0.553 (95% CI: 0.488–0.617); p > 0.05). Modified cut-off values for the QUS T-score improved the performance of CM-200 in identifying subjects with osteopenia (sensitivity 67.7% (95% CI: 62.8–72.3%); specificity 72.8% (95% CI: 68.1–77.2%); J = 0.405; AUC 0.702 (95% CI: 0.666–0.739); p < 0.001) and osteoporosis (sensitivity 79.4% (95% CI: 70.0–86.9%); specificity 61.8% (95% CI: 58.1–65.5%); J = 0.412; AUC 0.706 (95% CI: 0.654–0.758); p < 0.001). Conclusion: The modified cut-off values significantly improved the performance of CM-200 in identifying individuals with osteoporosis. Since these values are device-specific, optimization is necessary for accurate detection of individuals at risk for osteoporosis using QUS.

Highlights

  • The rapid ageing of the global population brings forth many noncommunicable diseases, including osteoporosis and its associated fragility fractures

  • A total of 910 subjects were recruited in the study, but 20 of them were excluded for receiving thiazide diuretics, 32 for glucocorticoids, 4 for cancer treatment, 30 for hormone treatment, 12 for having mobility problems, 5 for hysterectomy before menopause and 21 for not completing study procedure

  • The current study indicated that the agreement between quantitative ultrasound (QUS) and dual-energy X-ray absorptiometry (DXA) at the current cut-offs was significant but weak

Read more

Summary

Introduction

The rapid ageing of the global population brings forth many noncommunicable diseases, including osteoporosis and its associated fragility fractures. At QUS T-score ≤−2.5, CM-200 was ineffective in identifying subjects with osteoporosis (DXA T-score ≤−2.5) (sensitivity 14.4% (95% CI: 8.1–23.0%); specificity 96.1% (95% CI: 94.4–97.4%); AUC 0.553 (95% CI: 0.488–0.617); p > 0.05). Modified cut-off values for the QUS T-score improved the performance of CM-200 in identifying subjects with osteopenia (sensitivity 67.7% (95% CI: 62.8–72.3%); specificity 72.8% (95% CI: 68.1–77.2%); J = 0.405; AUC 0.702 (95% CI: 0.666–0.739); p < 0.001) and osteoporosis (sensitivity 79.4% (95% CI: 70.0–86.9%); specificity 61.8% (95% CI: 58.1–65.5%); J = 0.412; AUC 0.706 (95% CI: 0.654–0.758); p < 0.001). Conclusion: The modified cut-off values significantly improved the performance of CM-200 in identifying individuals with osteoporosis Since these values are device-specific, optimization is necessary for accurate detection of individuals at risk for osteoporosis using QUS

Methods
Results
Discussion
Conclusion