Abstract
Performance tests of a movable gamma-ray waste drum scanning system were carried out in this study. The scanner consists of a single HPGe detector on a movable cart with a 90° collimation angle to the observed item. The detection efficiency for 137Cs was determined using the ISOTOPIC software. The activities of 35 low-level radioactive waste drums were measured using this system and compared with the results of two other measurement systems. This movable scanning system gave the same result as the other two systems with a relative spread (one standard deviation) of about 7.2%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.