Abstract

Pacer-Cardioverter-Defibrillator. This article reviews the function of the pacer-cardioverter-defibrillator (PCD). Detection of ventricular arrhythmias occurs in two programmable zones, with onset and stability modifiers available to diminish overdetection of sinus tachycardia and atrial fibrillation, respectively. The sensing circuitry utilizes an auto-adjusting sensitivity with exponential decay to allow detection of low-amplitude ventricular fibrillation electrograms without T wave oversensing. Treatment can be accomplished by tiered therapy with two types of antitachycardia pacing, cardioversion and defibrillation. Cardioversion and defibrillation shocks are programmable between single pathway when two leads are used and simultaneous or sequential shock delivery when a three-lead system is used. A telemetered marker channel and electrogram aid in assessing device function during implantation and follow-up. Previously published literature is cited to expand on various aspects of PCD function and programming.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.