Abstract

The present study of 2,387 recordings of 12 anterior teeth in 53 patients indicates that the incisal edge is the optimal placement-site for the electric pulp tester to determine the lowest response threshold. The results show significant individual variations in the lowest threshold responses of the cervical-third, middle-third, incisal one-third, and incisal edge sites on a tooth, with a confidence level of 99%, according to the analysis of variance. The maxillary teeth gave a higher response threshold than the mandibular teeth and different types of teeth (canines and incisors) had statistically significant different response thresholds. The application of the electric pulp tester to the incisal-edge region with exposed dentin produced the most significant decrease in the threshold response.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.