Abstract
There are a number of parameters which have to be chosen depending on the analysis being done in gas chromatography. While the choice of stationary phase material is based on the solutes to be separated, the thickness is dependent on the concentration and the volatility of the components to be analyzed. This study undertakes a coupled column phase ratio optimization by connecting a short piece of a particular column prior to a normal length of an analytical column. Various columns of different dimensions (phase ratio), but of the same stationary phase material (methyl silicone), are coupled together by a deactivated glass press-fit connector, and the efficiency and capacity are measured. The coupling of fused silica open tubular columns is optimized in efficiency by matching or decreasing the phase ratio of successive columns. Capacity optimization is accomplished by increasing the phase ratio of consecutive columns. Capacity and efficiency optimization are opposing each other; therefore, if some efficiency can be sacrificed a substantial increase in capacity is possible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.