Abstract

ABSTRACT Objective Disintegrating cystine and calcium oxalate monohydrate stones present a formidable challenge owing to their hardness and distinct composition. This study aimed to establish optimal laser settings for these hard stones lithotripsy. Patients and Methods Cystine and calcium oxalate monohydrate stones were extracted from two patients. Two experiments were conducted in vitro by utilizing a 272 μm laser fiber with variable settings to disintegrate the cystine and calcium oxalate monohydrate stones. In the first experiment, energy was adjustable while frequency was constant, whereas the second experiment involved constant energy with adjustable frequency on each type of stone and each experiment was repeated three times to ensure robustness and reliability. Results Our findings indicated that for cystine stones, use of higher total power with high energy and low frequency proved to be effective. Conversely, for calcium oxalate monohydrate stones, settings involving higher total power with low energy and high frequency demonstrated superior efficacy and safety. Conclusion Holmium (Ho: YAG) laser settings with higher total power, high energy, and low frequency effectively disintegrate cystine stones despite increased heat, which was measured by a thermometer with a thermocouple. For calcium oxalate monohydrate stones, higher total power, high frequency, and low energy settings are recommended and safe.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.