Abstract
Newborn disease screening increases survival, improves quality of life and reduces treatment costs for healthcare systems. Mass spectrometry (MS) is an effective method for metabolic screening. However, conventional analytical methods require biofluid handling and cooling conditions during transport, making the logistics difficult and expensive, especially for remote regions. ’Paper-spray’ (PS) ionization generates a charged solvent spray from samples deposited on paper strips. Therefore, samples can be applied on a suitable matrix and shipped as dried spots to diagnostic laboratories with standard postal or messenger services. We built a robotic platform, the ’Open SprayBot’, to automatically analyze paper-deposited samples via PS-MS and increase the sample throughput. The system is operated via RUMBA32 and Arduino Mega boards. A commercial syringe pump and power supply provide solvent application and electrical current required for PS-MS. The usability of the Open SprayBot was demonstrated by quantifying palmitoyl-l-carnitine, a common biomarker in newborn screening.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.