Abstract

This paper presents the Outer Plug Standardization – OPS, a novel technique as an alternative for the classic internal standardization using multiple injection in capillary zone electrophoresis is proposed herein. This technique was applied in a new method for the determination of chloride, nitrate and sulfate in rainwater samples. After the injection accuracy was tested and proved to be a minor error source (average 1.26% RSD), the OPS was applied and it improved the intra-day and inter-day precision of the analytical method by 32.5% and 24.7%, respectively. Using a capillary with effective length of 23.5 cm the electrophoretic separation of the three inorganic anions could be achieved in <1 min, with detection limits of 0.05, 0.09 and 0.11 mg L−1 for chloride, nitrate and sulfate, respectively. Also, the statistical t-test was applied to the results obtained for 82 rainwater samples that were collected and analyzed applying both the method developed in this study and the official APHA 4140 method where no statistical difference was noted within a 95% confidence level.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.