The energy dependence of the response of single crystalline oxide scintillators (BGO, CWO and GSO(Ce)) was studied over the wide energy range from 6 to 1332 keV. The non-proportionality was defined as the difference between the specific response value at the given energy and the one at 661.6 keV. It is found that a considerable (up to 35%) non-proportionality of the response exists for the all investigated scintillators in the low energy region from 6 to 100 keV, while as in the energy range from 100 to 1332 keV the scintillator's response is proportional to the gamma-rays energy with the accuracy equal to /spl plusmn/1%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.