Abstract

BackgroundAn inhibitor can develop in congenital hemophilia A (HA) patients against exogenous infused factor (F)VIII, whereas in acquired HA (AHA) inhibitors initially develop against endogenous FVIII. Inhibitors can be detected with the Nijmegen Bethesda Assay (NBA), which has an international cut-off level of 0.60 Nijmegen Bethesda Units/mL (NBU/mL). Thereby, very low-titer inhibitors may remain undetected. AimTo describe the design and validation of the Nijmegen ultra-sensitive Bethesda Assay (NusBA) for the detection of very low-titer inhibitors. MethodsThe NusBA is a modification of the NBA in which the ratio of patient plasma to normal pooled plasma is changed from 1:1 to 9:1. Analytical validation was performed according to the CLSI EP10 guideline in order to determine trueness and reproducibility. Clinical validation was performed in two cohorts of congenital HA patients (82 adults) with pharmacokinetic data and four AHA patients. The limit of quantitation (LOQ) was determined by measuring plasma samples spiked with inhibitor levels in the low range (0.05–0.80 NBU/mL). ResultsThe LOQ for the NusBA was 0.10 NusBU/mL, with a coefficient of variation of 24.2 %. Seven (8.5 %) congenital HA patients had a positive NusBA result, of which only one was detected with the NBA. There was no correlation between NusBA and FVIII half-life. In three of the AHA patients the NusBA remained positive, when the NBA became negative. DiscussionThe NusBA is able to detect very low-titer FVIII inhibitors of ≥0.10 NBU/mL. Thereby, it may have added value in early inhibitor detection and therapy adjustments in patients with congenital HA and AHA.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call