Abstract
ABSTRACTAn accurate knowledge of the absolute calibration of the polarizing elements in the Near Infrared Camera and Multi‐Object Spectrometer (NICMOS) becomes especially important when conducting studies that require measuring degrees of polarization of close to 1% in the near‐infrared. We present a comprehensive study of all previously observed polarimetric standards using the NIC2 camera on NICMOS. Considering both pre‐ and post‐NICMOS Cooling System observations, we find variations in the polarimetry consistent with the effects of subpixel misalignments and the point‐spread function. We also measure nonzero results from unpolarized standards, indicating an instrumental polarization of p ≈ 1.2%, θ ≈ 88°. The lack of polarized and unpolarized standard stars with which to perform a comprehensive calibration study means that we cannot be confident that the current calibration will be effective for a number of recent large NICMOS GO programs. Further observations of polarimetric standards are needed in order to fully characterize the behavior of NICMOS at around p = 1%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Publications of the Astronomical Society of the Pacific
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.