Abstract

A thorough characterization of the key features of the new small-angle neutron scattering instrument SANS-1 at MLZ, a joint project of Technische Universität München and Helmholtz Zentrum Geesthacht, is presented. Measurements of the neutron beam profile, divergency and flux are given for various positions along the instrument including the sample position, and agree well with Monte Carlo simulations of SANS-1 using the program McStas. Secondly, the polarization option of SANS-1 is characterized for a broad wavelength band. A key feature of SANS-1 is the large accessible Q-range facilitated by the sideways movement of the detector. Particular attention is hence paid to the effects that arise due to large scattering angles on the detector where a standard cos3 solid angle correction is no longer applicable. Finally the performance of the instrument is characterized by a set of standard samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.