Abstract

Radiocarbon measurements on the carbonaceous aerosol fractions have been demonstrated as an effective tool for aerosol source apportionment. For these measurements, a new sample preparation facility was installed at the INFN-LABEC laboratory of Florence (Italy). The line was designed to allow the preparation of samples from different carbonaceous fractions: the combustion of the aerosol samples can be performed in helium or oxygen flows, according to thermal sequences. The evolved CO 2 is cryogenically trapped and reduced to graphite, which is the target material for following Accelerator Mass Spectrometry (AMS) 14C measurements. This preparation line is described in detail in the paper. As a first step, the line was tested by means of AMS measurements performed on standards to check the reproducibility and the accuracy of the system; moreover, preliminary measurements on the total carbon fraction in aerosol samples were made. Results of these measurements are also reported.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.