Abstract
A new water calorimeter for orthovolt x-rays has been developed at NMi. The purpose of this calorimeter is to provide calibrations in terms of absorbed dose to water. Four internationally accepted CCRI qualities (100–250 kV) have been characterized. Correction factors have been determined with measurements, Monte Carlo calculations and heat transport models. The results of the calculations have been validated against measurements. The absorbed dose to water calibration coefficient, NDw,WCM, with the new calorimeter has been compared to the NDw,NCS based on an air-kerma calibration coefficient, using the current NCS-10 dosimetry protocol. A good agreement is found for all beam qualities where the total uncertainty (1 SD) has decreased from 2.5% for NDw,NCS to better than 1.5% for NDw,WCM, using the new water calorimeter.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.