Abstract

The Cadmium‐Lined Outer‐Core Irradiation Tube (CLOCIT) is a new irradiation facility for 40Ar/39Ar geochronology at the Oregon State University TRIGA® reactor. We report fluence (i.e., time‐integrated flux) parameters from the first four CLOCIT irradiations and compare them with the existing Cadmium‐Lined Inner‐Core Irradiation Tube (CLICIT). CLOCIT provides an average neutron flux equivalent of 1.45–1.53 × 10−4 J h−1; about 55% of CLICIT. Radial fluence gradients were on the order of 0.2–4.2% cm−1. A planar fit of J‐values results in residuals in the range of uncertainty in the J‐value, but systematic deviations resolve a non‐planar component of the neutron flux field, which has also been observed in CLICIT. Axial neutron fluence gradients were 0.6–1% cm−1, compared with 0.7–1.6% cm−1 for the CLICIT. Production rate ratios of interfering reactions were (40Ar/39Ar)K = (4 ± 6) × 10−4 and (38Ar/39Ar)K = (1.208 ± 0.002) × 10−2, (36Ar/37Ar)Ca = (2.649 ± 0.014) × 10−4, (38Ar/37Ar)Ca = (3.33 ± 0.12) × 10−5 and (39Ar/37Ar)Ca = (9.1 ± 0.28) × 10−4, similar to the CLICIT values.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.