A new 250kV single stage AMS accelerator (SSAMS) was installed at the Center for Applied Isotope Studies, University of Georgia. The accelerator is intended to be used primarily for radiocarbon measurements of natural and biobased samples, while all other samples such as marine, geological, atmospheric and archaeological samples are measured on the decade-old 500kV compact tandem accelerator (CAMS). The new AMS system is equipped with a 134-cathode MC-SNICS ion source. In this article, we show the results of the tests carried out on standards and blanks and compare the performance of the new machine with that of the CAMS unit. We have also compared the stable isotope data from AMS measurements to the conventional isotope ratio mass spectrometers (IRMS) data.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.