Abstract

A directional neutron spectrometer named NCT-WES (Neutron Capture Therapy Wide Energy Spectrometer) was developed for quality assurance of the therapeutic neutron beam in Neutron Capture Therapy (NCT). NCT-WES operates as a “parallelized” Bonner spheres spectrometer, embedding six semiconductor-based thermal neutron detectors in a collimated cylindrical moderator. With the objective of validating the simulation model used to derive its response matrix, irradiations in reference monoenergetic fields were performed at National Physical Laboratory (UK). As the energy distributions of neutron beams in NCT extend from keV to a few MeV, monoenergetic fields in this domain were chosen, namely 71.5 keV, 144.2 keV, 565.1 keV, 841.9 keV and 1200.4 keV. The results of the experiment confirm the correctness of the NCT-WES simulation model, within an overall uncertainty lower than ± 2%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.