Abstract

The manganese salt bath is considered a primary standard for determining the absolute emission rate of radionuclide neutron sources. The National Research Council of Canada has recently revived its manganese salt bath and a full description of the system is given here. The physical characteristics of the bath, as well as the methods for determining the efficiency of the bath system and the induced activity in the bath, are described. An in-depth analysis of the fraction of neutrons captured in the manganese and the correction factor for neutron losses is also provided. Finally, the results of emission rate measurements of four different sources, complete with an uncertainty budget, are given. The emission rates of three americium-beryllium neutron sources and one californium-252 neutron source were found to agree with the known values, within a standard uncertainty of 1.7%.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call