Abstract

Calibration of the stiffness of atomic force microscope (AFM) cantilevers is critical for industry and academic research. The multi-position calibration method for AFM cantilevers based on vibration is investigated. The position providing minimum uncertainty is deduced. The validity of the multi-position approach is shown via theoretical and experimental means. We applied it to the recently developed vibration method using an AFM cantilever with a normal stiffness of 0.1 N m−1. The standard deviation of the measured stiffness is 0.002 N m−1 with a mean value of 0.189 N m−1 and the relative combined uncertainty is approximately 7%, which is better than the approach using the single position at the tip of the cantilever.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.