We aimed to find out whether the Hastings bipolar prosthesis moved in a bipolar fashion after its use for displaced fractures of the femoral neck. In 65 patients non-weight-bearing movement was assessed radiographically and compared with an earlier study of the Monk prosthesis. In 70% of patients, the only movement was between the acetabulum and the prosthetic shell, the prosthesis behaving as if it were unipolar. This did not change with time, but there was a slight improvement in the range of movement, particularly in flexion. The absence of movement between the outer shell and the metal head is most probably due to the design of the prosthesis and makes this prosthesis unsuitable for use in osteoarthritic hips.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.