Abstract
The responses of several types of surface contamination monitor have been investigated for three typical working surfaces contaminated by three radionuclides commonly used in diagnostic or analytical biomedical applications. Of the three radionuclides, is an intermediate-energy gamma emitter, decays by electron capture, emitting low-energy x-rays, and is a moderate-energy beta emitter. For each radionuclide monitored values have generally been found to be within 30% of projected response, this being based on calibrations which make use of standard radionuclide sources and conditions. The several types of contaminated non-absorbent surface that have been investigated produce monitored values which for a given type of monitor and radionuclide cannot be differentiated from each other.