Abstract
In this paper we describe measurements of the contrast transfer function, modulation transfer function and point-spread function of an optical coherence tomography (OCT) imaging system through scattering layers having a dimension-less scattering depth over the range 0.2–6.9. The results were found to be insensitive to scattering density, indicating that these measurement parameters alone do not well characterize the practical imaging ability of an OCT instrument. Attenuation and increased noise floor due to optical scattering were found to be the primary imaging limit and the effect of multiple scattering on OCT resolution was negligible.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.