Abstract
The authors have investigated the tongue-and-groove effect on the IMRT dose distributions using the EGS4/MCDOSE Monte Carlo code. They have compared the dose distributions computed by MCDOSE using the fluence maps with and without tongue-and-groove effect. Their results showed that, for one intensity modulated treatment field, the tongue-and-groove effect could be seen clearly in the dose distributions. For an IMRT treatment with multiple gantry angles, the difference between the dose distributions with and without the tongue-and-groove effect was reduced significantly. After considering the setup errors, however, no distinguished discrepancies were found, indicating that, for IMRT treatment planning, the tongue-and-groove effect can be neglected due to its negligible impact on the overall dose distributions.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.