Abstract
An osmolal gap (a disparity between measured and calculated serum osmolality) may accompany alcohol intoxication. We encountered a patient with methanol toxicity in whom no such gap was present, despite a markedly elevated serum methanol level. Further investigation revealed that serum osmolality had been measured with a vapor pressure osmometer; this technique may not detect volatile solutes such as alcohols. In vitro testing confirmed the insensitivity of vapor pressure osmometry to toxic ranges of both methanol and ethanol. When alcohol toxicity is suspected, an osmolal gap should be sought only if freezing point osmometry is available.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.