Abstract
This paper describes a simple microwave apparatus to measure the Hall effect in semiconductor wafers. The advantage of this technique is that it does not require contacts on the sample or the use of a resonant cavity. Our method consists of placing the semiconductor wafer into a slot cut in an X-band (8–12 GHz) waveguide series tee, injecting microwave power into the two opposite arms of the tee, and measuring the microwave output at the third arm. A magnetic field applied perpendicular to the wafer gives a microwave Hall signal that is linear in the magnetic field and which reverses phase when the magnetic field is reversed. The microwave Hall signal is proportional to the semiconductor mobility, which we compare for calibration purposes with d.c. mobility measurements obtained using the van der Pauw method. We obtain the resistivity by measuring the microwave reflection coefficient of the sample. This paper presents data for silicon and germanium samples doped with boron or phosphorus. The measured mobilities ranged from 270 to 3000 cm2/(V s).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.