Abstract

Periprosthetic fractures can be treated by various methods. The Mennen femoral plate used to be a common implant in our region to stabilise periprosthetic femoral fractures following hip arthroplasty. This device has been used in 16 patients in our region from three different centres. The periprosthetic femoral fractures occurred approximately 7 years after the hip arthroplasty procedure. After stabilisation of the fracture, patients were kept non-weight bearing. Weight bearing was commenced once there was clinical and radiological evidence of fracture union. About 75% of our patients had a complication following fixation of the fracture. In all of these patients the main complication was varus mal-union of the fracture. As per our study the Mennen femoral plate seems to be a weak fixation device. The plate is unable to counter the medial compressive forces on the femur leading to a varus collapse of the fracture.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.