Abstract
The object of this work was to evaluate the influence of the meniscus of liquid in the determination of viscosity by the oscillating-cup method. We measured the viscosity of water at 298.15 K with several heights of waterinside the cup. Results obtained allow us to state that, if the cup is not almost completely filled, errors in the viscosity measurements can be larger than the uncertainty of the method. But, if the cup is filled up to 95% of the total height, the error in the viscosity determination can be smaller than 0.1%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.