A model was developed which allows the dilatation of iliac cadaver arteries under normal blood pressure conditions (mean pressure 100 mmHg). Segments of the iliac artery were put under pressure and dilated with balloons under similar conditions as in clinical practice. Purposely oversized balloons were used because most of the arteries did not contain severely stenotic plaques and consequently the commercially available balloons had no effect because of their comparatively small diameter. In spite of the use of oversized balloons, the cadaver arteries could not be significant dilated. Rupture of the intima and media was almost invariably present. Leakage of distending fluid through the adventitia occurred commonly. The "therapeutic range" between minimal dilation and partial or total rupture in postmortem studies is small.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.