Abstract

The aim of this study was to evaluate the measurement precision and accuracy of T1 mapping using a polarity corrected (PC) TI prep tool, which was based on fast field echo (FFE) and obtained one data point with one inversion recovery (IR) pulse. A phantom was used consisting of eight materials with different Gd concentrations. T1 mappings were measured by changing the trigger interval and the inversion time (TI) interval. The T1 mapping measurement precision using the PC TI prep tool increased as the trigger interval was made longer. The measurement precision didn't depend on the interval of TI. On the other hand, when the trigger intervals are more than 1000 ms, the measurement accuracy was less than approximately 8%. By setting the optimal end of TI, the T1 mapping using a PC TI prep tool could measure the T1 value precisely and accurately.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.