Abstract

The objective of this paper is to present a new method of measuring oxygen diffusion coefficients with membrane-covered polarographic oxygen electrodes and to assess its applicability to viscous polymeric solutions of sodium carboxymethylcellulose (CMC), dextran and polyethylene glycol (PEG). The results indicate that, except in CMC solutions, the oxygen diffusion coefficient decreases with an increase in polymer concentration. In the CMC solutions, however, oxygen diffusion coefficients first decrease with an increased polymer concentration, then increase while the polymer concentration exceeds a critical value. Possible explanations for this phenomenon are explored.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.